Products
Search
Manufacturers
Services
Calibrations (Accredited laboratory) Maintenance - Repairs Department of Cleanroom Validation (Lloyd's Register Certified)) IQ/OQ/PQ Temperature/ Humidity Mapping Occupational Health & Safety Measurements Validation Studies Vibration and Noise studies Laboratory OrganizationECscan10L/M/H
TECHNICAL SPECIFICATIONS
|
Model |
ECscan10L |
ECscan10M |
ECscan10H |
|
Conductivity Range |
1.0~199.9uS/cm |
10~1999uS/cm |
0.1~19.99mS/cm |
|
Conductivity Accuracy |
±1% Full Scale |
||
|
Calibration Points |
1 point |
||
|
Calibration Solutions |
146.5uS/cm |
1413uS/cm |
12.88mS/cm |
|
Temperature Compensation |
0~50°C |
||
|
Temperature Coefficient |
2% per °C |
||
|
Cell Constant |
K=1 |
||
|
Normalization Temperature |
25°C |
||
|
Auto-Off |
8 minutes after last key pressed |
||
|
Sensor Type |
Platinum (Order Code: ECscan-C1-100K) |
Platinum (Order Code: ECscan-C10-100K) |
|
|
Connector |
6-pin |
||
|
Operating Temperature |
0~60°C, 32~140°F |
||
|
Power Requirements |
3×1.5V "G13A" Batteries |
||
|
Battery Life |
Approximately 150 hours |
||
|
Dimensions |
185(L)×40(Dia.)mm |
||
|
Weight |
100 grams
|
||