Electric Muffle Furnaces R Series

Μuffle furnace 3L or 8L up to 1100 °C with touch screen

Manufacturer : JP SELECTA

FOR TEMPERATURES ADJUSTABLE UP TO 1100 °C 
MICROPROCESSOR CONTROL WITH TFT TOUCH SCREEN
PRECISION ±2 °C OF THE SET VALUE
RESOLUTION: 1 DIGIT

FEATURES 


Metal external case with vent at the back of the unit. Interior and door made of ceramic fibre, resistant and durable (No asbestos). 
Heater situated at the side and bottom of the chamber. USB and RS-232 output. 


CONTROL PANEL 


General 
Main switch. TFT touch screen 4.3’’. Clock calendar. Two working modes, normal or programming. SPA – FRE – ENG menu. Self-test on starting. Temperature control auto-tuning. ºC/ºF selection. Type K probe. 
Normal mode 
Set point temperature selection Up ramp or no ramp. Stability time from 1 min to 99h or continuous. 
Programming mode 
10 programs capacity. 6 segments per program. Stability time in each segment from 1 min to 99h (or continuous in the last segment) Up ramps between segments or no ramps Daily - weekly On / Off programming.
 

Alarms 
Network failure detection alarm. Probe error detection alarm. Over temperature and low temperature alarms. Visual audible warning alarms. Up to 100 alarms storage (date, start time, end time and alarm type). 

 

Datalogging 

Datalogging memory up to 15000 data. Logging interval from 5 seconds to 30 min. Data download via RS -232 or USB. PC software for on-line registration (via RS-232). 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 