Calibrations (Accredited laboratory) Maintenance - Repairs Department of Cleanroom Validation (Lloyd's Register Certified)) IQ/OQ/PQ Temperature/ Humidity Mapping Occupational Health & Safety Measurements Validation Studies Vibration and Noise studies Laboratory OrganizationCertoClav EL
Benefits at a glance
- Compact: benchtop autoclave located directly at the laboratory workplace.
- Fast: short cycle times thanks to direct heating
- Practical: always processing the right quantity. No waiting for the big autoclaves
- Reliable: often more than 20 years in use. Lifelong service
Technical capability
- Power output: selectable between 125°/140°C. Special valve 115°/121°C available (0.7 - 2.7 bar / 10.2 - 39.2 psi overpressure). Solution of culture media (Agar) at 100°C
- Venting: venting using flow process. Steam saturation: non-condensable gases <3% (as mandated in EN 13060)
- Sterilisation material: solid instruments, liquids (culture media) bottles and devices, waste sterilisation
- Size, loading: optimal usage of chamber volume thanks to loading from above. Even long instruments up to 42 cm (18 L version) can be processed
Safety systems
- Tried-and-tested technology: 50 years of experience in autoclave construction
- Multiple systems: The 4-stage CertoSecure safety system consists of an excess pressure valve, pressure valve, safety cover locking device against opening to pressure and overpressure safety recess.
- Safe operation: the CertoLock lock closes safely with just one easy step. No need for time-consuming screwing as with other makes.
Type and fittings
- Controller: pressure control via valve, temperature control via thermostat.
- Instruments: precision manometer with drag pointer, precision thermometer
- Connections: No additional water connection required. Electrics: 220V, 50-60Hz, 1900W
- Standards: complies with international standards. GS mark in accordance with DIN and VDE/IEC (EN61010:1:2001 and 61010:2:040:2005). CE-conform.
- Heater: integrated in base. Heating elements are not in contact with water, therefore extended service life