Products
Search
Manufacturers
Services
Calibrations (Accredited laboratory) Maintenance - Repairs Department of Cleanroom Validation (Lloyd's Register Certified)) IQ/OQ/PQ Temperature/ Humidity Mapping Occupational Health & Safety Measurements Validation Studies Vibration and Noise studies Laboratory OrganizationDigicen 21R
TECHNICAL SPECIFICATIONS
The centrifuge refrigerated Digicen 21 R can reach up to 24.959 xg with a wide range of rotors:
- Swing out rotors, with a max volume of 4 x 100 ml which can fit several set of adapters
for tubes with or without caps, microtubes, microplates. - Angle fixed low speed for routine purposes.
- Angle fixed high speed from microtubes to 50 ml conical tubes.
On LCD screen shown the selectable values as time, RPM, RCF, temperature and braking time as well as graphics for lid open, progress bar, and acoustic signals for a better control about the process by operator.
The powerful refrigeration system, enhanced by the pre-cooling program, keep the samples at stable and controlled temperature, avoiding sample´s temperature fluctuations at introducing into the device.
RT 143![]() |
RT 138 (1)![]() |
RT 150![]() |
RT 183![]() |
RT 128![]() |
|
| + info | + info | + info | + info | + info | |
| ROTOR | SWING OUT | SWING OUT | SWING OUT | ANGLE FIXED | HORIZONTAL |
| Max. capacity | 4 x 50 ml. | 4 x 100 ml. | 2,4 o 6 Microtiter | 32 de 0,2 ml. | 24 de 1,5 x 75 ml. |
| RPM | 5300 | 5000 | 4000 | 16500 | 13000 |
| Radius(mm) | 145 | 147 | 122 (4) | 55 (4) | 86 |
| RCF Max. (xg) | 4554 | 4019 | 2182 | 16741 | 16250 |
| Min. temp. max speed (°C) | -4 | -4 | 0 | 2 | 1 |
RT 151 (3)![]() |
RT 110![]() |
RT 106![]() |
RT 108![]() |
RT 121 (3)![]() |
|
| + info | + info | + info | + info | + info | |
| ROTOR | ANGLE FIXED 45º | ANGLE FIXED 35º | ANGLE FIXED 35º | ANGLE FIXED 35º | ANGLE FIXED 45º |
| Max. capacity | 24 de 1,5-2 ml. | 24 x 5 ml. | 18 x 15 ml. | 24 x 15 ml. | 6 x 50 ml. |
| RPM | 16500 | 6500 | 4200 | 4200 | 6000 |
| Radius(mm) | 82 | 113 | 132 | 132/114 | 132 |
| RCF Max. (xg) | 24959 | 5338 | 2603 | 2603/2248 | 5313 |
| Min. temp. max speed (°C) | 2 | 1 | 0 | -1 | 1 |
RT 152 (3)![]() |
RT 153 (3)![]() |
RT 154 (3)![]() |
![]() |
![]() |
|
| + info | + info | + info | |||
| ROTOR | ANGLE FIXED 30º | ANGLE FIXED 30º | ANGLE FIXED 28º | ||
| Max. capacity | 12 x 10 ml Hermét | 8 x 30 ml Hermét | 6 x 50 ml Hermét | ||
| RPM | 15000 | 13500 | 9000 | ||
| Radius(mm) | 78 | 92 | 101 | ||
| RCF Max. (xg) | 19621 | 18746 | 9146 | ||
| Min. temp. max speed (°C) | 4 | 4 | 1 |
(1) This rotor includes hermetic lids.
(2) Fitting this tubes rotor lid can not be closed.
(3) Please, check tubes feature.
(4) Medium radio on bucket.