Products
Search
Manufacturers
Services
Calibrations (Accredited laboratory) Maintenance - Repairs Department of Cleanroom Validation (Lloyd's Register Certified)) IQ/OQ/PQ Temperature/ Humidity Mapping Occupational Health & Safety Measurements Validation Studies Vibration and Noise studies Laboratory OrganizationSteam distillation unit KJELDAHL PRO-NITRO-S
DETERMINATION OF ORGANIC NITROGEN (KJELDAHL METHOD)
AUTOMATIC BORACIC AND NaOH DOSAGE, SAMPLE DRAINAGE AND TEMPORIZED STOP
Semi-automatic steam distillation Kjeldahl unit. Simple secure systematic analysis suitable for medium to large throughput of samples.
FEATURES
- Steam distillation system.
- Compact steam generator with safety over temperature thermostat and over pressure device.
- Safety door, the system will not operate if the door is open.
- “Tube in place” sensor: if the tube is not located, the dosing process of NaOH will not take place.
- Universal adapter for digestion/distillation tubes MACRO (Ø 42mm) and MICRO (Ø 26 mm).
- Small footprint, saves bench top space: The H2O, NaOH and H3BO3 reservoirs are placed within the unit.
- Empty Digestion/Distillation tube system. Stainless steel case with reinforced ABS plastic front.
- Green LED 2 digit display.
- Distillation program: (Add NaOH, Add Boric Acid, Distillation time, Empty tube.)
- Automatic distillate titration kit. (See accessories).
SPECIFICATIONS
Measuring range: from 0.2 to 200 mg Nitrogen.
Programmable distillation time.
Nitrogen recovery >99.5%
Distillation speed : from 35-40ml/minute
Typical distillation time: from 7-10 minutes.
Water consumption rate: from 80-100 litres/Hr.
Steam generator water consumption: 2.5 Litres/ Hr.
Water reservoir for steam generator: 6 litres
NaOH reservoir: 2 Litres. Boric Acid reservoir: 2 Litres
ALARMS
Low water level for the steam generator.
Safety door open or no distillation/digestion tube in place.
Steam generator over temperature.
AUTOMATIC
Single push button to start the distillation cycle:
- Boric acid dosing
- Start distillation.
- NaOH dosing
- Stop Distillation (Programmed time transpired.)
- Acoustic indicator at the end of the cycle.
ADDITIONAL REQUIREMENTS
To complete Kjeldahl Nitrogen analysis a digestion block is also required.