Produits
Recherche
Fabricants
Services
Étalonnages (Laboratoire accrédité) Maintenance - Réparations IQ/OQ/PQ Validation/ Cartographie de la température et de l\'humidité Contrôles Santé et Sécurité Études de validation Études de Vibration et de Bruit Organisation de LaboratoireExtracteur de récupération des solvants Det-gras N (SOXHLET Methode)
Extracteur de récupération des solvants Det-gras N (SOXHLET Methode)
Modèles avec 2 ou 8 places
Fabricant : JP SELECTA
FOR THE DETERMINATION OF FATS BY SOXHLET METHOD. SAFETY PROTECTION IP65.
Equipment for the extraction of fats and soluble substances that will dissolve in solvents. Based on the Soxhlet methods by Randall, the DETGRAS-N extracts fats (and or other substances) from the sample by dissolving them in a solvent. If compared to the classic standard soxhlet method, a time saving of between 30-70% can be obtained due to a two stage extraction process and high speed in obtaining high temperatures.
APPLICATIONS
Practically all methods of classic soxhlet extraction can be substituted by the DET-GRAS N. Typical fat determinations are: in meat products, animal fodder, pre-prepared foods, fish etc. Also suitable for the extraction of soluble components in paper paste, textile fibres etc.
FEATURES
Heating by protected electric heater, distributed throughout the radiator surface that provides homogeneous heating throughout the samples.
Electrical safety according to IP65.
Robust temperature control by a durable thermocouple.
Safety over temperature device.
Outer case made of solvents resistant epoxy coated steel.
Extractions can be completed using glass or cellulose thimbles.
Compatible with most common solvents: Petroleum ether. Diethyl ether, hexane, acetone acetonitrile etc.
Supplied complete with 2 sets of seals, adapters for different solvent types.
Typical extraction time (food fats) 50 minutes.
SPECIFICATIONS
Cellulose extraction thimbles Øext 26 x 60 mm.
Reusable glass thimbles Øext 34 x 80 mm.
Working temperature of 90 to 240 °C.
Solvent recovery 60 to 80 %. Solvent volume (by sample): up to 50 ml.
Program memories: 16. Extraction time <>: from 0 to 99 minutes.
Extraction time <>: from 0 to 99 minutes.
Time to recuperate solvents: from 0 to 99 minutes.
CONTROL PANEL
LCD display of 2 lines of 20 characters.
Keypad with 4 push buttons to select temperature, time and programs.